Beam Profiler Measurements Explained: Centroid, Width, Ellipticity and Peak

IZAK Scientific OEM board-level laser beam profiler module, front view

In short

  • Every result answers one question. The centroid says where the beam is, the widths say how big it is, ellipticity and azimuth say what shape it has, and the peak value says whether the image can be trusted.
  • Check the peak first. A saturated or very weak image makes every other number wrong, however many decimals it shows.
  • Background is the silent error. An offset of 1% of full scale that is not subtracted can make a second-moment width several times too large and drag the centroid toward the middle of the frame.
  • Report the definition with the number. A width without its method (second moment or threshold), area of interest and background method cannot be compared with another instrument.

What a beam profiler reports

A camera beam profiler records an image of the beam and turns it into a short list of numbers. The list looks similar in most profiler software, but each number is a different calculation with its own weak points. The table below is the reading key for the rest of this guide.

ResultWhat it isWhat it tells youMost sensitive to
Centroid (X, Y)Intensity-weighted centre of the beamBeam position, alignment, pointing driftBackground offset, beam clipped at the sensor edge
Peak value and positionBrightest pixel and where it isWhether exposure is right; hot spotsNoise (one pixel), saturation
Major and minor widthBeam size along the two principal axesSpot size, focus, collimationWidth definition, background, saturation
AzimuthAngle of the major axisOrientation of an elliptical beamMeaningless for round beams
Ellipticity, eccentricityMinor ÷ major width, and √(1 − ellipticity²)How round the beam isPixel size on small beams, background
Sum of countsTotal of all pixel values in the analysis areaRelative power at fixed settingsExposure, gain, attenuation (not a calibrated power)

Centroid and beam position

The centroid is the centre of mass of the intensity: every pixel value I is multiplied by its coordinate and the total is divided by the sum of all pixel values.

Xc = Σ(I · x) ÷ ΣI     Yc = Σ(I · y) ÷ ΣI

Because it averages thousands of pixels, the centroid is far more stable than the position of the peak pixel and can resolve movements much smaller than one pixel. That is why it is the right quantity for alignment and for pointing stability measurements.

Two things bias it. A background offset that is not subtracted adds a uniform “beam” the size of the whole frame, whose centre is the centre of the image, so the result moves toward the middle of the frame. In the explorer below, a 1% offset moves the centroid of a 30-pixel beam by about 6 pixels. A beam that is cut by the sensor edge loses part of its weight, so its centroid moves inward. In an asymmetric beam the centroid and the peak are not at the same place, and that difference is real information about the beam, not an error.

Beam width: second moment and threshold

Profiler software reports the width along the beam’s major and minor axes. How that width is calculated matters more than any other setting. The two families you will meet are:

  • Second-moment width (D4σ). Four times the standard deviation of the intensity distribution, the reference definition of ISO 11146. It is defined for any beam shape and is the width used for M² and divergence, but every pixel in the analysis area contributes, so background and noise in the wings affect it strongly.
  • Threshold (clip-level) widths. The extent of the region above a chosen level, for example 13.5% of the peak (the 1/e² level of a Gaussian beam). They ignore everything below the threshold, so they are robust against background, but they change with the beam shape and the threshold chosen.

For an ideal Gaussian beam the two agree. For other shapes they do not: in the explorer, a flat-top beam 30 pixels across has a D4σ of only 27.5 pixels, and a Gaussian with a 5% pedestal has a threshold width of 32 pixels but a D4σ of about 36. Neither is wrong; they measure different things. The full comparison of D4σ, 1/e², FWHM, knife-edge and power-content widths is in Beam width definitions.

Major axis, minor axis and azimuth

A real beam is rarely perfectly round, so the software describes it by an ellipse. The principal axes come from the second moments of the image, σx², σy² and σxy, and the azimuth is the angle of the major axis:

θ = ½ · atan[2σxy ÷ (σx² − σy²)], reported between 0° and 180°

Always quote widths along the principal axes, not along the camera rows and columns: a beam rotated by 45° looks round when it is measured along X and Y. When the beam is nearly round, the azimuth is defined by noise and can jump from one frame to the next; ignore it when the ellipticity is above about 0.95. The difference between an elliptical beam and an astigmatic beam, which needs measurements at several planes, is explained in Beam astigmatism and ellipticity.

Ellipticity and eccentricity

Ellipticity is the ratio of the minor to the major width, d ÷ D. It is 1 for a round beam and falls toward 0 for a line. ISO 11146 treats a beam as round when its ellipticity is greater than 0.87. Eccentricity, √(1 − (d ÷ D)²), expresses the same shape on the opposite scale: 0 for a circle, close to 1 for a line.

Ellipticity d ÷ DEccentricityTypical meaning
1.000.00Round beam
0.950.31Round within measurement noise for most beams
0.870.49Limit for “round” according to ISO 11146
0.700.71Clearly elliptical, for example an uncorrected module
0.300.95Diode fast and slow axis without circularization

Some programs report ellipticity as D ÷ d or as 1 − d ÷ D. Check which definition a datasheet uses before you compare numbers. On small beams the ellipticity also depends on the pixel size: with only a few pixels across the minor axis, one pixel more or less changes the ratio by several percent. See small vs large pixels for the sampling rule.

Peak value, saturation and the sum of counts

The peak value is the brightest pixel in the analysis area, in camera counts or as a percentage of full scale. Read it before anything else:

  • At full scale the top of the beam is cut off. The true peak is hidden and the widths come out too large: in the explorer, a beam driven to twice full scale gives a D4σ 7% too large and a 13.5% threshold width 19% too large.
  • Below about 20% of full scale the wings sink into noise and quantization, and widths become unstable.
  • Between roughly 50% and 90% the image uses the dynamic range of the camera without clipping. Adjust exposure, gain or attenuation to get there.

The peak position is the location of a single pixel, so it is noisy and, on a saturated beam, it is simply the first clipped pixel the software finds. Use the centroid for position and the peak position only to locate hot spots. The sum of all counts in the analysis area is proportional to beam power only while exposure, gain and attenuation stay the same. It is a useful relative indicator for drift, not a calibrated power reading; use a power meter for watts.

Tool

Beam results explorer

Build a beam on a simulated 160 × 120 pixel, 8-bit camera and see how each result reacts. Change the background, the exposure or the area of interest and compare the numbers with the true beam.

Examples:
Beam position on the sensor
D4σ ellipse and centroidISO area of interestpeak pixel · red = saturated

Model: each pixel = background + peak × I(x, y), rounded to 8 bits (0–255) and clipped at 255. Gaussian I = exp(−2r²), flat-top I = exp(−2r¹⁰), pedestal adds 0.05·exp(−r²/2), with r the elliptical radius normalized to the 1/e² (or edge) half-widths. Second moments are taken about the centroid; D4σ = 2√2·[σx² + σy² ± √((σx² − σy²)² + 4σxy²)]½. The ISO area starts from the pixels above 13.5% of the peak and is then iterated as an ellipse with axes 3 × D4σ until it stops changing. Threshold widths are the extent of the pixels above 13.5% of the peak along the principal axes, in whole pixels. “True” values are the second moments of the noise-free beam without background, clipping or sensor edges. The simulation illustrates the definitions; it is not the algorithm of any particular software.

Area of interest and background: the hidden settings

Two settings that are often left at their defaults decide whether a second-moment width means anything.

Background. Block the laser, record a dark frame with the same exposure and gain, and subtract it, or let the software estimate the offset from a beam-free part of the image. In the explorer, a 1% offset that is not subtracted makes the D4σ of a 30-pixel beam more than four times too large and turns its azimuth from 30° to about 2°, while the 13.5% threshold width hardly moves. With the offset subtracted, the error falls to about 2%.

Area of interest. ISO 11146 limits the calculation to an area about three times the beam width, centred on the beam, so that distant noise cannot add to the second moments. The explorer shows the limit of this rule: without background subtraction, the iterative area keeps growing and ends up at the whole frame. The area of interest reduces noise from far pixels; it does not replace a dark frame.

Keep the whole beam, wings included, inside both the sensor and the area of interest. A beam cut by the sensor edge gives a smaller width, an inward-shifted centroid and a false ellipticity.

Turning results into pass/fail limits

On a production line the same results become acceptance limits. A practical set for a collimated laser module could look like this:

CheckExample limitWhy
Peak value20–90% of full scaleRejects images that cannot be measured before any other test
Centroid positionWithin ±50 µm of the referenceOptical axis and mounting alignment
Major and minor widthNominal ±5%, same definition every timeCollimation and focus
Ellipticity≥ 0.87Round beam according to ISO 11146
Sum of countsWithin ±10% of a golden unit, fixed settingsGross power drop or contamination

Set each limit from the measurement repeatability first: run one unit 30 times, and make the tolerance several times larger than the scatter. How to build the full test, including golden units and gauge studies, is in beam profiler production QA.

What to record with every result

  • Width definition (second moment or threshold, and the threshold level) and whether widths are along the principal axes.
  • Background method and area of interest.
  • Peak value as a percentage of full scale, exposure, gain and attenuation.
  • Pixel size and any magnification of imaging optics used for the µm calibration.
  • Measurement plane: distance from the laser aperture or from the focus.
  • Number of frames averaged, and for pulsed lasers, pulses per frame.

IZAK products for this application

IZAK laser beam profilers report the centroid and beam position, the major and minor beam width, azimuth, ellipticity and eccentricity, and the peak value and position, with a fixed or automatic area of interest. A Production Test Mode applies saved acceptance limits and gives a one-click pass/fail, and results are saved as CSV files, images and reports.

Longer wavelengths: MWIR 1.5–5.7 µm and LWIR 7.7–12 µm. New to beam profiling? Start with the complete guide to laser beam profiling.

What to send us for a recommendation

With these details an engineer can propose a profiler, the optics and the settings for your measurement:

  1. Laser type and wavelengthCW or pulsed, pulse width and repetition rate
  2. Beam size at the measurement planeand whether it is focused, collimated or diverging
  3. Power or energyaverage and peak
  4. Results you needposition, width, ellipticity, pass/fail limits
  5. Width definition in your specificationsecond moment, 1/e² or another level
  6. Where it will be usedlab, incoming inspection or production line
Not sure your results mean what you think?Send us a raw beam image and the numbers your software reports. We will tell you which settings to check.

Frequently asked questions

QWhat is the centroid of a laser beam?
It is the intensity-weighted centre of the beam: the sum of each pixel value times its coordinate, divided by the sum of all pixel values. It is more stable than the peak position and is used for alignment and pointing measurements.
QWhy is my D4σ beam width much larger than expected?
Usually because of background. A small offset that is not subtracted adds weight far from the beam, and the second moment grows quickly with distance. Subtract a dark frame, limit the area of interest to about three beam widths and check that the image is not saturated.
QWhat is the difference between ellipticity and eccentricity?
Ellipticity is the minor-to-major width ratio, 1 for a round beam. Eccentricity is √(1 − ellipticity²), 0 for a round beam. They describe the same shape on opposite scales. ISO 11146 treats a beam with ellipticity above 0.87 as round.
QWhat peak level should a beam profiler image have?
Aim for roughly 50–90% of full scale. At full scale the top of the beam is clipped and widths come out too large; below about 20% the wings are lost in noise.
QCan a beam profiler measure laser power?
The sum of counts is proportional to power only while exposure, gain and attenuation stay constant, so it works as a relative indicator for drift. For absolute power or energy, use a calibrated power or energy meter.
QWhy do two beam profilers give different widths for the same beam?
They often use different width definitions, thresholds, background methods or areas of interest. Compare results only with the same definition, the same calibration and the same background treatment.

References

  1. ISO 11146-1: Lasers and laser-related equipment — Test methods for laser beam widths, divergence angles and beam propagation ratios — Part 1: Stigmatic and simple astigmatic beams.
  2. ISO 11146-3: Lasers and laser-related equipment — Test methods for laser beam widths, divergence angles and beam propagation ratios — Part 3: Intrinsic and geometrical laser beam classification, propagation and details of test methods.
  3. ISO 13694: Optics and optical instruments — Lasers and laser-related equipment — Test methods for laser beam power (energy) density distribution.
>Tzachi Sabati Tzachi Sabati
CEO, IZAK Scientific
Physicist specializing in photonics and quantum technologies, with deep expertise in quantum sensors and advanced optical systems. Leads the Advanced Quantum Lab course at the Technion, bridging academic excellence with industry innovation. At IZAK Scientific, provides cutting-edge photonics-based solutions, developing customized inspection and sensing systems for R&D and production. Passionate about advancing quantum sensing applications and integrating novel technologies to meet industry needs.

 

Get new IZAK guides and tools by email

One short email a month: new technical articles, free calculators and product updates in laser measurement and electro-optics. No spam, unsubscribe anytime.

Get new IZAK guides and free calculators by email, one short email a month: Subscribe · Privacy Policy
Skip to content