In short
- Test the fiber, not only the laser. The delivery fiber is the part of a surgical laser that changes most between treatments. Its transmission, end faces and tip decide what reaches the tissue.
- Measure transmission by substitution. Read the power at the launch plane, then at the distal tip with a detector that catches the whole output cone, and divide.
- Know the loss budget. Two uncoated silica faces cost about 6 %, a few metres of low-OH glass about 1 %. A larger gap points to coupling, end-face or bend loss.
- Inspect both ends and the output beam. Microscope the proximal and distal faces, check the cleave angle, and look at the far field for halos and rings.
Surgical laser fiber testing answers a simple question: of the power the laser console reports, how much actually leaves the fiber tip, and in what kind of beam? Holmium (about 2.1 µm) and thulium fiber lasers (about 1.94 µm) for lithotripsy and soft tissue, Nd:YAG at 1064 nm and diode lasers in the near IR all deliver their energy through a silica fiber with a core of a few hundred micrometres. That fiber is connectorised, bent through an endoscope, pressed against stones or tissue, and often replaced or reprocessed after each case. This guide sets out a practical test plan for fiber makers, medical laser manufacturers and service teams: what to measure, how to measure it, what the numbers should look like, and how to tell a bad fiber from a bad measurement.
What surgical laser fiber testing has to prove
A medical laser is specified at the working end. IEC 60601-2-22 requires the delivered output to stay within 20 % of the value the device indicates, and the output must be checkable with a calibrated meter. Every loss between the laser cavity and the fiber tip eats into that margin, and the fiber is where most of the variation sits: different lots, different connector polish, different users, different numbers of uses.
A fiber that fails rarely fails in only one way. The common failure modes are:
- Low transmission from a poor proximal face, a misaligned launch or excess bend loss. The patient receives less energy than the screen shows.
- Proximal connector damage. When the launch spot or cone overfills the core, light enters the cladding and the ferrule, heats the connector and can burn the input face.
- Distal tip damage. In contact lithotripsy the tip erodes and shortens (burn-back) and the face fractures, which changes both power and beam shape.
- Breakage or leakage in a bend. A deflected endoscope bends the fiber hard near the tip. Light lost in the bend heats the jacket and can damage the scope.
The tests below catch each of these. The same plan works for incoming inspection of purchased fibers, final test in fiber assembly, design verification of a new fiber, and service checks on a returned system. For how fiber tests fit into the wider verification plan of a medical laser, see our guide to verification and validation of a laser medical device.
Six tests that catch most fiber failures
Distal power divided by launch power at the treatment wavelength, measured with a detector that captures the full output cone.
Microscope image of the input face: scratches, chips, burn marks, epoxy and contamination, plus fiber centring in the connector.
Flatness and angle of the output face, chips, hackle and cracks, and the length and condition of the stripped tip.
Far-field image of the output: divergence, NA fill, symmetry, and halos or rings that reveal cladding light or a tilted launch.
Transmission straight and wrapped on a mandrel at the bend radius of the endoscope, with a check for hot spots along the jacket.
Transmission, length and face image before and after a defined firing protocol, for design verification and for reprocessed fibers.
Measuring transmission: the substitution method
The cut-back method used for telecom fibers is destructive and does not fit a finished, connectorised surgical fiber. For assemblies the practical method is substitution, sometimes called insertion loss:
- Reference. Couple the test source into a short, well-characterised reference fiber of the same core and NA, or measure directly at the launch plane, and record Pref.
- Test. Connect the fiber under test with the same connector seat and record the distal power Pout.
- Result. Transmission T = Pout / Pref. Report it in percent and, if your specification uses decibels, as −10·log10(T).
Three details decide whether that number means anything:
- Catch the whole cone. A fiber with NA 0.22 emits into a full angle of 2·arcsin(0.22) ≈ 25.4°. At 10 mm from a 365 µm core the output is already about 4.9 mm across. Put the tip close to a detector with a large aperture, or use an integrating sphere, so that no light misses the sensor.
- Test at the treatment wavelength. Fresnel reflection, glass absorption and bend loss all depend on wavelength. A visible aiming beam is useful for a quick continuity check, not for an acceptance value.
- Reproduce the launch. The laser’s own coupling optics set the spot size and cone angle at the input face. A test bench that launches a smaller, slower cone will make every fiber look better than it will perform in the system.
Power detectors for this work must handle the full power at the tip, or a known fraction of it, at 1.9–2.1 µm or wherever your laser runs. Thermal detectors are the usual choice because their response is nearly flat across the near and mid IR. Our article on the multi-channel power meter for high-power lasers shows how fiber-coupled channels with SMA inputs let you measure several fibers or outputs at once.
The transmission budget: what a good fiber should give
Before you set a pass limit, work out what physics allows. For a clean, well-launched fiber, three terms dominate.
Fresnel reflection at each uncoated face. R = ((n − 1)/(n + 1))². Fused silica has n ≈ 1.436 at 2.12 µm, so R ≈ 3.2 % per face and two faces pass 93.7 %. Anti-reflection-coated or index-matched faces remove most of this.
Absorption in the glass. Holmium and thulium wavelengths sit close to a strong OH absorption band near 2.2 µm, so fibers for these lasers use low-OH silica; high-OH silica is the choice for the UV. Over the few metres of a surgical fiber, even 20 dB/km costs only 0.05 dB over 2.5 m, about 1.1 %.
Coupling at the input. Keep the focused spot inside about 70 % of the core diameter and the launch cone inside the fiber NA. Light launched outside either limit goes into the cladding, where it is lost in the first bend or heats the connector. Our fiber coupling efficiency guide covers the launch optics in detail.
| Loss term | Typical size | How to see it | Fix |
|---|---|---|---|
| Fresnel, two uncoated faces | about 6.3 % at 2.1 µm | Always present; constant from fiber to fiber | AR coating, or include it in the limit |
| Glass absorption, low-OH silica, 2–3 m | about 1 % (at 20 dB/km) | Grows with length | Correct glass type for the wavelength |
| Overfilled launch | a few % to tens of % | Halo in the far field, hot connector | Smaller spot, slower cone, better alignment |
| Damaged or dirty end face | a few % up to total loss | Microscope image, burn mark | Clean, repolish or re-cleave, or reject |
| Bend loss | negligible above the critical radius, steep below it | Transmission falls when wrapped on a mandrel | Bend-tolerant fiber design, larger radius |
Worked example: a 365 µm holmium fiber
A 2.5 m fiber, 365 µm core, NA 0.22, two uncoated faces, low-OH silica at 20 dB/km, used with a holmium laser at 2.12 µm. The launch spot is 240 µm with a launch NA of 0.15. On the bench, Pref = 20.0 W and Pout = 17.6 W.
- Expected transmission: 0.9369 × 0.9886 = 92.6 %.
- Measured transmission: 17.6 / 20.0 = 88.0 %.
- Excess loss: −10·log10(0.880/0.926) = 0.22 dB.
- Launch: the spot fills 66 % of the core and the cone uses 68 % of the NA, both inside the guidelines.
With an 85 % acceptance limit the fiber passes. But the launch is clean and the glass explains only about 1 %, so 4.6 percentage points are unexplained. The next step is a microscope look at both faces and a bend check, before this fiber lot is released. If this happens across a lot, the cause is usually a process issue such as polishing, cleaving or connector bonding rather than the individual fibers.
Surgical fiber transmission and launch check
Enter the fiber, the launch and your two power readings. The calculator gives the transmission you should expect from Fresnel reflection and glass absorption, the measured transmission and the unexplained excess loss, checks whether the launch overfills the core or the NA, and shows how far a cleave angle tilts the output beam.
Model: refractive index of fused silica from its Sellmeier equation; Fresnel loss per uncoated face R = ((n − 1)/(n + 1))² at normal incidence; glass transmission 10−αL/10; excess loss = −10·log10(Tmeasured/Texpected). Launch guidelines: spot ≤ about 70 % of the core, launch NA ≤ fiber NA. Exit tilt from a cleave angle θ in air: arcsin(n·sin θ) − θ. The core of a real fiber is doped and the attenuation depends on the glass, so use your fiber data sheet values where you have them. Bend loss, connector offset and face damage are not modelled: they show up as excess loss.
Inspecting the end faces and the cleave angle
The proximal face carries the highest irradiance in the system: tens of watts in a spot of a few hundred micrometres. Any particle, residue or scratch there absorbs, heats and grows. Image the face under a microscope at a magnification that resolves features of a few micrometres, with both bright-field and oblique illumination, and look for burn marks, pits, chips at the core–cladding edge, epoxy on the face and fiber recession or protrusion in the ferrule. The zone-based acceptance approach of our fiber end-face inspection guide adapts well: define a core zone where no defect is allowed and an outer zone with size limits.
The distal face is usually cleaved or polished flat. Look for hackle, lips, chips and cracks, and check that the stripped length and any coating or cap are as specified. A tilted face tilts the beam. Refraction at the exit face turns a cleave angle θ into a beam tilt of arcsin(n·sin θ) − θ in air: 0.44° for a 1° cleave at 2.12 µm, and about 1.35° for a 3° cleave at 1064 nm. In saline the tilt is much smaller (about 0.2° for a 2° cleave), because the index step is smaller. Angle matters more for non-contact use and for aiming than for contact lithotripsy, so set the limit from the application, not from habit.
Checking the output beam with a beam profiler
Power tells you how much light leaves the tip. A far-field image tells you how it leaves. Point the attenuated output of the fiber at a beam profiler sensor, or onto a diffuse screen imaged by the profiler, at a known distance, and measure the second-moment diameter at two distances. The full divergence angle is 2·arctan((D2 − D1)/(2Δz)), and the effective output NA is the sine of half that angle. Our divergence measurement guide gives the procedure.
What to look for:
- NA fill much larger than the launch NA means mode mixing in tight bends or a damaged face scattering light into high angles.
- A bright halo or sharp outer ring is cladding light: the launch overfills the core or the connector is off-centre.
- A ring (donut) pattern means the beam enters the fiber at an angle to its axis, a typical sign of launch tilt.
- An off-axis centroid on a straight fiber is the cleave angle at work.
At 1.94 µm the IZAK SWIR beam profiler (up to 2.05 µm) and the MWIR beam profiler (1.5–5.7 µm) both apply. For holmium at 2.1 µm, use the MWIR family. For Nd:YAG and diode lasers up to 1100 nm, the UV–VIS–NIR profiler (300–1100 nm) covers the wavelength. Always attenuate: a camera sensor needs only a tiny fraction of a surgical laser’s output.
Bend loss and tip degradation
Bend loss is negligible above a critical radius and rises steeply below it, so a fiber can pass a straight test and fail inside a deflected endoscope. Measure transmission with the fiber straight, then wrapped around a mandrel at the tightest radius of the intended scope, with the deflected section placed near the tip as in use. Watch the jacket with a thermal camera or a thermocouple during the test: a local hot spot shows where light leaves the core long before the jacket fails.
Tip degradation is a design-verification and reprocessing question. In contact lithotripsy the tip erodes and shortens with use, and the rate depends on pulse energy, pulse duration and the target. To compare fiber designs, fire a defined protocol into a standard target and record, before and after: transmission, tip length against a reference mark, and a microscope image of the face. For reprocessed fibers, re-strip and re-cleave to the specified length, then repeat the full transmission and face inspection as for a new fiber.
A test plan that fits production
| Stage | Tests | Sample |
|---|---|---|
| Design verification | All six tests, bend at worst-case radius, tip degradation protocol, repeat after sterilisation cycles if reusable | Several fibers from more than one lot |
| Final test in fiber assembly | Transmission at the treatment wavelength, proximal and distal face images, length and labelling | Every fiber |
| Incoming inspection | Transmission and proximal face image | Per lot, by sampling plan |
| Service and reprocessing | Transmission, both faces, far-field check | Every returned or reprocessed fiber |
Log every result with the fiber serial or lot number, the wavelength, the reference reading and the face images. Transmission trends across lots are often the earliest sign of a polishing or bonding drift in fiber production. For high volumes, a test station with fiber-coupled power channels, automated face imaging and a pass/fail report per fiber saves far more operator time than it costs.
IZAK products for this application
IZAK builds measurement systems for fiber and high-power laser delivery: multi-channel power measurement with fiber-coupled inputs, automated end-face inspection with pass/fail per fiber, and beam profilers from UV to LWIR for the output beam.
Power measurementMulti-Channel Power MeterFive fiber-coupled channels with SMA inputs, up to 50 W per channel, 200 nm to 20 µm, with parallel real-time readings and logging.View product →
Fiber inspectionFiber Ferrule Inspection SystemImages each ferrule end face and returns pass/fail for defects and cleanliness, center offset, adhesive symmetry and transmittance, as an application or API.View product →
UV to LWIRLaser Beam ProfilersUV–VIS–NIR 300–1100 nm, SWIR up to 2.05 µm, MWIR 1.5–5.7 µm and LWIR 7.7–12 µm, for far-field and divergence checks of the fiber output.View product →Need a complete fiber test station for production or service, with your connector, wavelength and report format? See our fiber optic design services or book an inspection consultation.
Frequently asked questions
QHow do you measure the transmission of a surgical laser fiber?
QWhy do holmium and thulium laser fibers use low-OH silica?
QWhat transmission should a good surgical fiber reach?
References
- IEC 60601-2-22, Medical electrical equipment — Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.
- IEC 60825-1, Safety of laser products — Part 1: Equipment classification and requirements.
- ISO 11554, Optics and photonics — Lasers and laser-related equipment — Test methods for laser beam power, energy and temporal characteristics.
- ISO 11146-1, Lasers and laser-related equipment — Test methods for laser beam widths, divergence angles and beam propagation ratios.
- IEC 60793-1-40, Optical fibres — Part 1-40: Attenuation measurement methods.
Tzachi Sabati
CEO, IZAK Scientific
Physicist specializing in photonics and quantum technologies, with deep expertise in quantum sensors and advanced optical systems. Leads the Advanced Quantum Lab course at the Technion, bridging academic excellence with industry innovation. At IZAK Scientific, provides cutting-edge photonics-based solutions, developing customized inspection and sensing systems for R&D and production. Passionate about advancing quantum sensing applications and integrating novel technologies to meet industry needs.