SWIR Laser Beam Profiling: 1064, 1310, 1550 nm and 2 µm Lasers

SWIR Laser Beam profiler for the 900nm -1700nm spectral range

In short

  • SWIR starts where visible-light sensors fade. Above about 1.1 µm you need a SWIR profiler. IZAK’s SWIR family covers 400–2050 nm in six models.
  • Check both ends of the band. At 1030–1064 nm both the UV–VIS–NIR and the SWIR families work. Above 2.05 µm, for example holmium at 2.1 µm, use the MWIR family.
  • Plan the pixels. SWIR sensors have fewer pixels than visible ones. Keep at least 10 pixels across the beam and the beam under one third of the sensor width; image small fiber spots with a relay lens.
  • Treat “eye-safe” with care. SWIR beams are invisible and still hazardous at high power. Align with viewer cards and the live profile, not by eye.

What SWIR means in beam profiling

Short-wave infrared (SWIR) is the band from about 0.9 to 2 µm. In beam profiling it begins where the sensors used for visible and near-infrared light lose their response, around 1.1 µm, and it ends where mid-wave infrared detectors take over. Telecom lasers, fiber lasers, eye-safe rangefinders and lidar, and medical thulium lasers all sit in this band, so it is one of the most common places to need a profiler that a standard camera cannot replace.

IZAK’s SWIR laser beam profilers cover 400–2050 nm: four models for 400–1760 nm, an extended model to 2050 nm and a fast NIR–SWIR model for 900–1700 nm. They use the same software as the rest of the family: live 2D and 3D profiles, centroid, widths, ellipticity, pass/fail tests and reports. For the basics of beam profiling, start with The Complete Guide to Laser Beam Profiling.

SWIR lasers and the profiler for each

The table lists the lasers most often profiled in the SWIR band and the IZAK family that covers each wavelength.

LaserWavelengthTypical useIZAK family
Pump diodes915–980 nmPumping fiber and solid-state lasersUV–VIS–NIR or SWIR
Ytterbium fiber and disk, Nd:YAG1030–1080 nmCutting, welding, marking, researchUV–VIS–NIR or SWIR
Telecom O-band1260–1360 nmData-center and access opticsSWIR
Raman and erbium pumps1450–1480 nmOptical amplifiersSWIR
Erbium fiber, telecom C and L bands1530–1625 nmTelecom, lidar, sensing, eye-safe rangefindersSWIR; MWIR from 1.5 µm
Thulium fiber1.9–2.05 µmMedical (strong water absorption near 1.94 µm), plastics processing, pumpingExtended SWIR (32005) or MWIR
Holmium2.05–2.1 µmMedical, mid-IR pumpingMWIR above 2.05 µm

The SWIR family ends at 2.05 µm and the MWIR family starts at 1.5 µm, so 1.5–2.05 µm is covered by both. Choose by beam size, frame rate and budget, or ask us.

What is different about profiling in SWIR

01You cannot see the beam

Alignment by eye is impossible. Use an IR viewer card or a viewer to find the beam, then centre it on the live profile at low power.

02Fewer, larger pixels

SWIR sensors usually have lower pixel counts than visible ones. Small beams need a model with more pixels or a relay lens.

03More dark signal

SWIR detectors produce more dark signal, and it changes with temperature and exposure. Take a dark frame at the same settings and subtract it.

04The 1 µm overlap

Visible-range profilers still respond at 1030–1064 nm, but their sensitivity falls steeply toward 1100 nm and the light spreads deeper in the sensor.

05Fiber outputs and small spots

A single-mode fiber at 1550 nm has a mode field about 10 µm wide, too small to image directly. Magnify it onto the sensor.

06High power, invisible

Fiber lasers in this band reach kilowatts. Sample and attenuate before the sensor, and plan the beam dump for light you cannot see.

Around 1 µm: UV–VIS–NIR or SWIR?

For ytterbium and Nd:YAG lasers at 1030–1064 nm both families can work. A UV–VIS–NIR profiler has more and smaller pixels, which helps with small beams, and only a small part of the laser power is needed even where the sensor is less sensitive. Its limits are signal near 1100 nm and a slight blur: at these wavelengths the light penetrates deep into the sensor before it is absorbed, and the charge spreads to neighbouring pixels. This matters most for small or sharp-edged beams. Above about 1.1 µm, pushing a visible-range sensor with more power gives blooming and wrong widths, so use a SWIR profiler. Our article on small versus large pixels explains how pixel size affects the measured width.

Tool

SWIR profiler finder

Enter the wavelength and the beam diameter at the sensor. The finder lists the SWIR models that cover the wavelength and checks two rules: at least 10 pixels across the beam, and the beam no wider than one third of the sensor, so there is room for the background area that ISO 11146 needs.

Not sure of the beam size? Use the Gaussian Beam Calculator to find the diameter at your measurement plane.

Small beams and fiber outputs

Many SWIR measurements are of small spots: the mode field of a fiber, a focused spot or the output of a waveguide. A standard single-mode fiber has a mode field diameter of about 9 µm at 1310 nm and about 10 µm at 1550 nm. With pixels a few micrometres wide that is only two or three pixels, which is not enough to measure a width.

The answer is relay imaging: a lens or microscope objective images the spot onto the sensor with a known magnification M. The profiler measures the magnified width, and the real width is that value divided by M. Three rules keep the result right:

  • Choose M for 10 or more pixels across the spot, while keeping the magnified beam under one third of the sensor width. For a 10 µm mode field and pixels of a few micrometres, that is about 4× to 10×.
  • Calibrate M, for example by moving a target a known distance or imaging a scale, instead of trusting the nominal value.
  • Use optics that do not clip the beam. The lens aperture must collect the full divergence of the fiber output, or the measured spot looks smaller and more Gaussian than it is.

The finder above gives the approximate magnification when the beam is under-sampled. For collimated beams of a few millimetres, the opposite applies: if the beam is wider than one third of the sensor, reduce it with a telescope or choose a model with a larger sensor.

Power, attenuation and safety

A SWIR profiler, like any camera sensor, needs only a tiny part of the beam: typically microwatts to milliwatts spread over the spot, depending on the model and the exposure. Continuous fiber lasers of hundreds of watts or kilowatts need a sampling stage first, a wedge or a pick-off that sends a small, known fraction of the beam to the profiler, then attenuators close to the sensor. Check that the attenuators are made for the wavelength: a filter that is neutral in the visible may not be neutral at 1550 nm or 2 µm. For pulsed SWIR sources, plan exposure and attenuation from the pulse energy, as explained in How to Profile a Pulsed Laser Beam.

Wavelengths above about 1.4 µm are absorbed in the front of the eye and do not reach the retina, which is why lasers near 1.5 µm are called eye-safe. That is relative: the allowed exposure is higher, not unlimited, and high-power SWIR beams still burn skin and eyes. Classify the laser under IEC 60825-1, wear eyewear rated for the wavelength, and remember that a stray reflection you cannot see is still there.

Choosing among the SWIR models

Six configurations cover the band. All are delivered as packaged systems with the IZAK software; the same detectors can also be integrated as an OEM module in your system.

ModelSpectral rangeSensor widthResolutionBest for
32001 SWIR Compact400–1760 nm2.5 mm0.3 MPSmall collimated beams and integration where space is tight
32002 SWIR Standard400–1760 nm5.2 mm1.3 MPGeneral lab work at 1 to 1.7 µm
32003 SWIR High Resolution400–1760 nm5.3 mm3.2 MPSmall spots, fiber outputs with relay optics, beam detail
32004 SWIR Large-Area400–1760 nm7.1 mm5.2 MPLarger beams with high detail
32005 Extended SWIR Fast400–2050 nm10.2 mm0.3 MPThulium lasers to 2.05 µm and large beams
32006 NIR-SWIR Fast900–1700 nm10.2 mm0.3 MPLarge telecom and 1.5 µm beams

Frame rate and interface are set for each order according to the optical power, beam diameter and integration needs. The SWIR product page has the full configuration of each model.

Checklist: profiling a SWIR laser

  1. Wavelengthinside the model range; above 2.05 µm use MWIR
  2. Sampling10 or more pixels across the beam, or relay optics
  3. Sensor sizebeam under one third of the sensor width
  4. Dark frametaken at the same exposure and temperature, subtracted
  5. Attenuationsampling first; filters rated at the laser wavelength
  6. AlignmentIR card or viewer, then the live profile at low power
  7. Safetyeyewear for the wavelength, beam dump for invisible light
  8. Reportwidth definition (D4σ, 1/e²) and magnification stated

Free PDF · 6 pages

Take the selection rules with you

  • Wavelength-to-detector map, 0.3 to 12 µm
  • Pixel-pitch and sensor-size rules with worked examples
  • Pulsed vs CW, attenuation and production QA checklists
  • A fill-in specification worksheet for your RFQ
Get the guide

IZAK products for this application

Profilers for SWIR lasers, from the lab bench to integration in your system.

Need a SWIR test station that is not in the catalog? We design custom electro-optical systems →

What to send us for a recommendation

With these details an engineer can recommend the model, optics and attenuation in one reply:

  1. Laser type and wavelengthe.g. erbium fiber laser, 1550 nm
  2. Power or pulse energyCW power, or energy, width and repetition rate
  3. Beam sizeat the planned measurement plane, or fiber type
  4. What you need to measurewidth, M², pointing, mode quality
  5. InterfaceUSB3 or GigE, cable length, software or API
  6. EnvironmentR&D bench, OEM integration or production line
Profiling a SWIR laser?Describe your laser and beam. We reply with a model, optics and attenuation proposal.

Frequently asked questions

QWhat wavelength range does a SWIR beam profiler cover?
IZAK’s SWIR profilers cover 400–2050 nm across six models: four for 400–1760 nm, an extended model to 2050 nm and a fast model for 900–1700 nm. Above 2.05 µm, use the MWIR family (1.5–5.7 µm).
QCan I profile a 1550 nm laser with a standard camera?
Not reliably. Sensors made for visible and near-infrared light lose their response around 1.1 µm. Pushing them with more power causes blooming and wrong beam widths. Use a SWIR profiler for 1310 and 1550 nm lasers.
QWhich profiler should I use for a 1064 nm laser?
Both the UV–VIS–NIR and the SWIR families cover 1064 nm. The UV–VIS–NIR models have more and smaller pixels, which helps with small beams; the SWIR models have more signal at this wavelength. For small or sharp-edged beams, ask us to compare.
QHow do I measure the mode field of a single-mode fiber?
Image the fiber end onto the sensor with a lens or objective of known magnification, usually 4× to 10× for a 10 µm mode field. Calibrate the magnification, make sure the lens collects the full divergence, and divide the measured width by the magnification.
QIs a 2 µm thulium laser SWIR or MWIR?
Both families reach it. The Extended SWIR model covers up to 2050 nm, and the MWIR family covers 1.5–5.7 µm. Holmium lasers near 2.1 µm are beyond the SWIR range and need the MWIR family.
QAre “eye-safe” SWIR lasers safe to align by eye?
No. Above about 1.4 µm the eye absorbs the light before the retina, so the allowed exposure is higher, but high-power beams still damage the eye and skin, and you cannot see them. Use eyewear rated for the wavelength and align with an IR card and the live profile.

References

  1. ISO 11146-1: Lasers and laser-related equipment — Test methods for laser beam widths, divergence angles and beam propagation ratios — Part 1: Stigmatic and simple astigmatic beams.
  2. ISO/TR 11146-3: Part 3: Intrinsic and geometrical laser beam classification, propagation and details of test methods.
  3. IEC 60825-1: Safety of laser products — Part 1: Equipment classification and requirements.
  4. ITU-T G.652: Characteristics of a single-mode optical fibre and cable (mode field diameter).
>Tzachi Sabati Tzachi Sabati
CEO, IZAK Scientific
Physicist specializing in photonics and quantum technologies, with deep expertise in quantum sensors and advanced optical systems. Leads the Advanced Quantum Lab course at the Technion, bridging academic excellence with industry innovation. At IZAK Scientific, provides cutting-edge photonics-based solutions, developing customized inspection and sensing systems for R&D and production. Passionate about advancing quantum sensing applications and integrating novel technologies to meet industry needs.

 