Common LED Production Test Mistakes: Binning, Temperature and Settling Time

Green indicator LEDs lit on a printed circuit board, seen through the imager of an IZAK LED tester during a production test

In short

  • The LED heats itself during the test. In every LED production test, flux, dominant wavelength and forward voltage all move with junction temperature, so the moment you measure is part of the result.
  • Pick a defined thermal state. Either a short pulse with the junction at the fixture temperature, or true steady state. A reading taken “a second or two after switch-on” is neither, and it is not repeatable.
  • Know your settling time. A small indicator LED on a board can need half a minute to settle within 2 °C, while the first 20 ms see almost no heating.
  • Guard-band the bins. With a 5 nm wavelength bin and ±1 nm measurement uncertainty, only the middle 3 nm can be accepted with confidence.
  • Measure the voltage at the LED. Contact and cable resistance in a two-wire set-up can hide or fake a forward-voltage shift as large as the thermal one.

Why LED production test results drift

An LED production test usually checks three things for every part or board: light output (flux or intensity), colour (dominant wavelength, peak wavelength or chromaticity) and forward voltage at a set current. The results feed bins, pass/fail decisions and traceability records. The trouble is that all three quantities depend on the temperature of the LED junction, and the junction heats up from the moment the test current is switched on.

Only part of the electrical power leaves the LED as light. The rest, typically more than half, is heat generated in the die. That heat flows through the package, the solder joint, the board and the fixture, and each stage has its own thermal resistance and its own time constant. The die itself warms within microseconds to milliseconds, the package within tenths of a second, and a board in a test fixture within tens of seconds or minutes. A single LED therefore has no single “warm-up time”: the reading you get depends on where on this heating curve you sample.

The effect is not small. Red and amber LEDs typically lose around 1 % of their light per °C of junction temperature and shift towards longer wavelengths by roughly 0.1 nm per °C. Blue, green and white LEDs are less sensitive in flux and wavelength, but their forward voltage still falls by about 2 to 3 mV per °C. Between a cold pulse test and the steady state of the final product, a red indicator LED can easily read 10–15 % brighter and more than 1 nm shorter in wavelength. That is enough to move a part across a bin boundary.

LED type (typical order of magnitude)Flux vs TjDominant wavelength vs TjForward voltage vs Tj
Red / amber (AlInGaP)about −1 %/°Cabout +0.1 nm/°Cabout −1.5 to −2 mV/°C
Blue / green (InGaN)about −0.1 to −0.3 %/°Cabout +0.02 to +0.05 nm/°Cabout −2 to −3 mV/°C
Phosphor-converted whiteabout −0.1 to −0.3 %/°Cchromaticity shift, check u′v′about −2 to −3 mV/°C

These are orders of magnitude for planning only. Use the temperature coefficients in your LED datasheet, or measure them on your own fixture, before you set limits.

Eight common LED production test mistakes

1. Measuring at an undefined point on the heating curve

Symptom: two testers, or the same tester before and after a software update, disagree by a few percent on the same parts. Cause: the spectrometer or photodetector is triggered at a time that is not fixed relative to the current step: a software delay, a variable USB latency, or an integration time that changed when the gain was adjusted. Each moves the reading along the heating curve. Fix: trigger the optical measurement in hardware from the current source, record the delay and the integration window in the test report, and keep the whole integration window inside the current pulse.

2. Binning cold and selling hot

Symptom: parts that passed in production look dimmer or shifted in the customer’s product. Cause: a short pulse test measures the LED with the junction at the fixture temperature, usually about 25 °C, while the product runs the LED with its junction tens of degrees hotter. Neither number is wrong, but they answer different questions. Fix: decide which condition the bin represents and write it into the specification: “single pulse at Tj = 25 °C”, or “DC at a stated case or solder-point temperature”. If the application runs hot, either test on a temperature-controlled heat sink at the application temperature, or apply a correction derived from measured coefficients and verify it on samples.

3. Not waiting for steady state, or waiting for the wrong one

Symptom: DC measurements creep during the test and depend on the takt time of the line. Cause: the test is meant to be at steady state but is taken after one or two seconds, when only the fast package stage has settled and the board and fixture are still warming. Fix: measure the settling curve once on your actual fixture, define a tolerance (for example “Tj within 2 °C of final”), and set the delay from it. If that delay is longer than the takt time allows, switch to a pulsed method rather than shortening the wait.

4. Letting the fixture warm up over the shift

Symptom: a slow trend in flux or wavelength over the first hour of production, repeated every morning. Cause: the fixture, nest or heat sink accumulates heat from thousands of test cycles, so the starting temperature of each LED rises. Fix: log the fixture temperature with every result, control it if the drift matters, and measure a reference LED (a golden sample) at the start of each shift and at fixed intervals.

5. Binning without a guard band

Symptom: customers find parts outside the bin, although every part passed. Cause: parts measured close to a bin edge are assigned with no allowance for measurement uncertainty. With an expanded uncertainty U (k = 2) of 1 nm, a part measured 0.5 nm inside the edge has a real chance of being outside it. Fix: set acceptance limits inside the bin limits by U, as in the ISO 14253-1 decision rules, and route parts in the guard zones to a retest or to a neighbouring bin. Then work on reducing U, because the guard zone is lost yield.

6. Two-wire forward voltage

Symptom: forward voltage varies from nest to nest and grows as the pogo pins wear. Cause: the voltage is measured at the source, so it includes the drop across cables and contacts. At 350 mA, 50 mΩ of contact resistance adds 17.5 mV, the same as a junction temperature change of about 9 °C at −2 mV/°C. Fix: use four-wire (Kelvin) sensing up to the LED pads. This matters even more if forward voltage is used to estimate junction temperature, as in the electrical test method of JESD51-51.

7. A current source that overshoots or rises slowly

Symptom: pulsed results depend on the pulse length in a way the thermal model cannot explain, or short-pulse results are noisy. Cause: an overshoot at switch-on heats the die more than the nominal current would, and a slow rise eats into a short pulse. Fix: look at the LED current on an oscilloscope once per fixture design. The rise time should be a small fraction of the pulse length, with no overshoot, and the current must be settled before the optical integration starts.

8. Spectrometer and geometry errors that look like LED variation

Symptom: bin populations shift after a spectrometer exchange or a change of integration time. Cause: saturated or near-dark spectra, missing dark subtraction, ambient light reaching the input, a wavelength calibration that has drifted, or a geometry that captures a different fraction of the light for parts with different beam patterns. Fix: keep the peak signal in the linear part of the range, take a dark spectrum at the same integration time, shield the measurement from room light, verify the wavelength scale with a line source on a schedule, and use a geometry that matches the quantity you bin: total flux in an integrating sphere, intensity at a defined distance and solid angle. Our comparison of the integrating sphere and the cosine corrector explains which geometry measures what.

Calculator

LED thermal drift and settling-time calculator

Enter the drive point, a simple two-stage thermal model of the LED on your fixture and the datasheet temperature coefficients. The calculator shows how far the junction has heated at the moment you measure, how much flux, wavelength and forward voltage differ from steady state, how long the LED needs to settle, and how much of the bin survives the guard band.

Model: heat Ph = I·Vf·(1 − light share); junction rise ΔTj(t) = Ph·[R1(1 − e−t/τ1) + R2(1 − e−t/τ2)], steady state Ph·(R1 + R2). Differences are reading at time t minus steady state: flux ratio ek·ΔT − 1, wavelength and voltage linear in ΔT. Settling time is when the remaining rise falls below your tolerance. Guard-banded acceptance zone = bin width − 2U. The presets are illustrative orders of magnitude, not data for a specific part: use your datasheet and, ideally, a measured heating curve of your fixture.

Worked example: a red indicator LED at 20 mA

A red indicator LED on a control board is driven at 20 mA with a forward voltage of 2.0 V, so it takes 40 mW. With about 15 % of that leaving as light, 34 mW turns into heat. Assume a thermal resistance of 250 K/W from junction to board with a time constant of 0.3 s, and 150 K/W from board to ambient through the fixture with a time constant of 30 s. The datasheet coefficients are −1 %/°C for flux, +0.1 nm/°C for dominant wavelength and −1.5 mV/°C for forward voltage. The wavelength bin is 5 nm wide.

Steady state. The junction ends 0.034 W × (250 + 150) K/W = 13.6 °C above the fixture.

Pulse test at 20 ms. The junction has risen only 0.55 °C. Compared with steady state, the LED reads 13.9 % brighter, its dominant wavelength reads 1.30 nm shorter and its forward voltage 19.6 mV higher. The wavelength offset is 26 % of the bin width.

Test at 2 s. The package stage has settled but the board has not: the junction is 8.8 °C up and still rising. The reading is 4.9 % brighter and 0.48 nm shorter than steady state, and it changes with every small variation in delay. This is the least repeatable point to test.

Steady state within 2 °C. The remaining rise drops below 2 °C after about 28 s. If the line cannot wait that long, use the short pulse and a documented correction, not a compromise delay.

Guard band. If the expanded uncertainty of the dominant wavelength measurement is ±1.0 nm, the acceptance zone of the 5 nm bin is 3 nm, 60 % of the bin. Parts measured within 1 nm of either edge go to retest or to the neighbouring bin. The calculator above reproduces all of these numbers with the “Red indicator, 20 mA” preset.

Which test mode answers which question

Test modeJunction stateGood forWatch out for
Single short pulse (milliseconds)Close to fixture temperatureFast production binning; comparing parts with each otherHardware trigger, current rise time, integration window inside the pulse
Continuous short pulses at low duty cycleClose to fixture temperatureCharacterisation at a controlled junction temperatureSynchronising the detector with the pulses; average versus peak values
DC after settlingSelf-heated, defined by the fixtureResults close to real operation on that fixtureLong settling time; fixture drift; result depends on the fixture
DC on a temperature-controlled heat sinkSet by the heat sink and the LED’s own heatingDatasheet-style values at a stated temperature, such as 85 °CThermal contact, waiting time, cost per station

Standards bodies distinguish these cases for exactly this reason. IES LM-85 describes single-pulse, continuous-pulse and DC operation for LED packages, and CIE 225 defines measurement conditions for high-power LEDs at a stated junction temperature. Whatever mode you choose, name it in the test specification together with the current, the timing and the fixture or heat-sink temperature.

An LED production test checklist

  1. State the thermal condition of every limit: pulse length and delay, or settling criterion, and the fixture or heat-sink temperature.
  2. Trigger the optical measurement in hardware from the current source and log the actual timing.
  3. Measure the settling curve once per fixture design and keep it with the test specification.
  4. Use four-wire sensing for forward voltage and check the current waveform once per fixture.
  5. Record the fixture temperature with every result.
  6. Measure a golden sample at the start of each shift and at fixed intervals, and stop the line if it moves outside its control limits.
  7. Set acceptance limits inside the bin limits by the expanded measurement uncertainty.
  8. Keep the spectrometer in its linear range, subtract a dark spectrum and verify the wavelength scale on a schedule.

For white LEDs, the same rules apply to chromaticity instead of dominant wavelength. Our CCT and Duv calculator converts chromaticity coordinates into the values most white-LED bins use, and the chromaticity quadrangles of ANSI C78.377 are the usual starting point for white bins. If your test checks indicator LEDs on assembled boards rather than single parts, see how to embed an LED checker in your production line.

IZAK products for this application

IZAK builds LED test solutions for production lines: an imaging tester that checks the colour, position and presence of every LED on a board, and custom radiometers and spectroradiometers for spectral and radiometric measurements. Test timing, current control and fixtures are designed together with the optics.

Need a complete LED test station with a pulsed current source, hardware triggering and fixture temperature logging? See our control and test automation engineering or book a radiometry consultation.

Is your LED test drifting?Send us the LED type, test current, timing, fixture and the bins you use. We will tell you where the thermal and measurement errors are likely to come from and how to fix them.

Frequently asked questions

QShould LEDs be binned with a pulse test or at steady state?
Either can be right, as long as the condition is defined and repeatable. A short single pulse measures the LED at the fixture temperature and is fast enough for production. Steady-state DC on a temperature-controlled heat sink gives values closer to real operation. Avoid readings taken partway up the heating curve, such as one or two seconds after switch-on.
QHow much does LED colour change with temperature?
Red and amber LEDs typically shift by about +0.1 nm per °C in dominant wavelength and lose about 1 % of their flux per °C. Blue and green LEDs shift less, about +0.02 to +0.05 nm per °C. A red indicator LED that heats by 13.6 °C therefore moves by about 1.3 nm, a quarter of a 5 nm bin. Use your datasheet coefficients for exact values.
QWhat guard band should I use for LED binning?
Set the acceptance limits inside each bin edge by the expanded measurement uncertainty U (k = 2) of your tester. With a 5 nm bin and U = 1 nm, accept parts measured in the middle 3 nm and send parts within 1 nm of an edge to a retest or the neighbouring bin.

References

  1. ANSI/IES LM-85, Measurements of LED sources: electrical and photometric measurement in single-pulse, continuous-pulse and DC operation.
  2. CIE 225:2017, Optical Measurement of High-Power LEDs.
  3. CIE 127:2007, Measurement of LEDs.
  4. ANSI C78.377, Specifications for the Chromaticity of Solid State Lighting Products.
  5. JESD51-51, Implementation of the Electrical Test Method for the Measurement of the Real Thermal Resistance and Impedance of Light-emitting Diodes with Exposed Cooling Surface.
  6. ISO 14253-1:2017, Geometrical product specifications (GPS) — Inspection by measurement of workpieces and measuring equipment — Part 1: Decision rules for verifying conformity or nonconformity with specifications.
>Tzachi Sabati Tzachi Sabati
CEO, IZAK Scientific
Physicist specializing in photonics and quantum technologies, with deep expertise in quantum sensors and advanced optical systems. Leads the Advanced Quantum Lab course at the Technion, bridging academic excellence with industry innovation. At IZAK Scientific, provides cutting-edge photonics-based solutions, developing customized inspection and sensing systems for R&D and production. Passionate about advancing quantum sensing applications and integrating novel technologies to meet industry needs.

 