In short
- An integrating sphere collects light emitted in every direction and spreads it evenly over its diffuse, highly reflective wall, so a detector at one port sees a signal proportional to the total flux of the source.
- It suits LEDs well: it captures the whole emission pattern and gives total flux, spectral power distribution and colour temperature from one set-up.
- Size it on purpose: keep the ports under about 5% of the sphere’s wall area, choose the coating for your wavelength range, and remember that a bigger sphere gives less signal and a slower response.
- IZAK built a bare-LED inspection system around a sphere with three channels (total power, spectrum and temporal pulse shape), custom optomechanics and software with pass/fail reports and pn-junction temperature estimation.
Introduction to the integrating sphere
An integrating sphere is a key tool in optical measurement, especially for characterising light sources such as light-emitting diodes (LEDs). Its design and function rest on the core principles of optics and photometry. This article covers how an integrating sphere is designed, how it works and the role it plays in LED inspection, and then describes an LED inspection system that IZAK Scientific built around one.
Design and working principle of the integrating sphere
An integrating sphere is a hollow spherical cavity whose inner surface carries a highly reflective, diffuse coating. Two coating materials are common:
Reflectivity above 98% across the visible range, with near-Lambertian (perfectly diffuse) reflection. A practical, lower-cost coating.
Reflectivity above 99% over much of the visible and near-IR, also near-Lambertian. Sintered PTFE is durable and easy to clean.
Openings for the light to enter and leave, and for mounting detectors and sources. They are placed so that no detector sees the source directly; the light must first reflect from the wall many times.
Small diffuse screens between the source and the detector ports block the first, direct reflection, so the detector sees only light that has already been spread over the wall.
Working principle
Light from a source, for example an LED, strikes the inner surface and is scattered in all directions. After many reflections the light is spread evenly over the whole wall. This homogenisation removes the spatial and angular variation of the source, so a detector placed at a port measures a uniform light level that is proportional to the total power entering the sphere.
The physics behind this is Lambert’s cosine law. A perfectly diffuse (Lambertian) surface reflects light with an intensity that follows
I(θ) = I₀ · cos θ
where I₀ is the intensity along the surface normal and θ is the viewing angle measured from the normal. Because the projected area of the surface shrinks by the same cos θ, the reflected radiance is the same in every direction. Inside a sphere this has a useful consequence: every point on the wall receives the same share of the light reflected from every other point, so after a few reflections the wall irradiance is uniform, whatever the original beam looked like.
The many reflections also multiply the light level. The wall radiance is higher than a single diffuse reflection would give by the sphere multiplier M = ρ / (1 − ρ(1 − f)), where ρ is the wall reflectance and f is the fraction of the wall taken up by ports. With ρ = 0.98 and 3% ports, M is about 19. The multiplier also explains why port area and coating quality matter so much: both directly change the signal and its stability. The tool below calculates it for your sphere.
Measurements with an integrating sphere
An integrating sphere supports several radiometric and photometric measurements:
The light emitted from, or passing through, an area within a given solid angle: L = d²Φ / (dA · dΩ · cos θ), in W/(m²·sr).
The radiant power incident on a surface per unit area: E = dΦ / dA, in W/m². At the sphere wall, E = π · L.
The radiant power emitted in a given direction per unit solid angle: I = dΦ / dΩ, in W/sr.
The temperature of a blackbody whose colour matches the source. For LEDs, which are not blackbodies, the correlated colour temperature (CCT) of the closest blackbody colour is used.
The radiant power per unit wavelength, in W/nm. Total flux, colour coordinates, CCT and luminous flux are all calculated from it.
For an introduction to these quantities and to the difference between radiometric (watts) and photometric (lumens) units, see The Basics of Light Measurement: Radiometry and Photometry.
Characterising light sources
Integrating spheres characterise light sources by measuring their total luminous flux, spectral power distribution and colour temperature. Because the sphere captures light emitted in all directions, it gives accurate and complete data, which is exactly what LEDs need: their emission pattern is wide and differs from part to part, so a single detector in front of the LED would measure only one slice of it.
Two geometries are used. A bare LED or a lamp that emits in all directions is placed at the centre of the sphere (4π geometry). A forward-emitting LED on a board can be mounted at a wall port facing into the sphere (2π geometry). In both cases the sphere is calibrated with a reference source of known total flux, and an auxiliary lamp can correct for the light that the device and its holder absorb inside the sphere.
Applications
Integrating spheres are used with many light sources, including LEDs, lasers and incandescent bulbs. They are essential in industries that need precise light measurement, such as display technology, automotive lighting and architectural lighting. With a laser, the sphere also acts as a diffuser that makes the power reading independent of beam size, position and polarisation.
Integrating sphere sizing and signal calculator
Enter the sphere diameter, the wall reflectance and the port diameters. The tool returns the port fraction, the sphere multiplier, the irradiance at the detector port, the power on your detector and the sphere’s time constant, which limits how fast a pulse it can follow. Set an unused port to 0.
Model: port fraction f = Σd²/(4D²); multiplier M = ρ/(1 − ρ(1 − f)); wall irradiance E = Φ·M/(πD²); power on the detector P = E·A; time constant τ = (2D/3c)/(−ln(ρ(1 − f))), from the mean path 2D/3 between reflections. Baffles, the device under test and coating ageing lower the real signal, so calibrate the sphere with a reference source.
Case study: integrating sphere for bare LED measurement by IZAK Scientific
IZAK Scientific developed an inspection system for bare LEDs based on an integrating sphere. The system has several measurement channels, each one dedicated to a different property of the LED:
Captures all the light the LED emits to determine its total radiant flux. A calibrated precision power meter gives an accurate reading; the luminous flux follows from the measured spectrum.
A high-resolution spectrometer measures the spectral power distribution, colour properties and colour temperature. It is calibrated for the spectral reflectance of the sphere.
A fast photodiode and an oscilloscope capture rapid changes in the light output, the LED’s performance over time and its response to pulsed drive currents at different repetition rates.

The temporal channel is where sphere size matters most. Light bounces inside the sphere for a few nanoseconds before it reaches the detector, so the sphere acts as a low-pass filter with a time constant of roughly 2 ns for a 2-inch sphere and around 7 ns for a 100 mm sphere with a high-reflectance coating. For LED pulses of microseconds this is negligible; for nanosecond pulses, choose a small sphere, or check the effect with the calculator above.
Optomechanical design
IZAK Scientific also supplied the optomechanical design that holds the LED board in position inside the sphere, for consistent and reliable measurements. Adjustable mounts place the LED at the centre of the sphere and align it with the measurement channels, so every unit is measured in the same geometry and results can be compared from part to part and from batch to batch.
Software application for the LED inspection system
The software supports all the acquisition channels: spectrum, power meter and temporal shape. It calculates and displays the number of photons per wavelength, which is the quantity that matters for photochemical, biological and detector-calibration applications.
The software logs every measurement result and generates inspection reports against predefined pass/fail criteria, such as spectrum width, energy per pulse and degradation over time. It is highly customisable: thresholds can be tailored and new measurement capabilities added to meet each customer’s requirements.

The software also calculates the pn-junction temperature, a key indicator of the performance and lifetime of laser diodes that is equally relevant to LEDs: junction temperature drives efficiency, wavelength shift and lifespan. Figure 4 shows a pulse sequence and the laser diode’s pn-junction temperature during operation while the assembly cools. This information helps predict how the device will perform and how long it will last.

The combination of measurement capability and reporting makes the software a valuable tool for checking the quality and reliability of LEDs and laser diodes. For a high-volume version of LED testing, see LED Inspection Part 2: LED and Shape Inspection for High-Volume Production, and for pulsed-lamp measurements, Xenon Flash Lamps in Classical and Quantum Innovations.
Conclusion
Integrating spheres are essential for LED inspection: they offer a complete way to measure the optical properties of LEDs, from total flux to spectrum, colour and temporal behaviour. IZAK Scientific provides tailored solutions, including customised integrating spheres. Contact us to select and customise the right integrating sphere for your needs.
IZAK Scientific can help you select, design and customise an inspection system, with or without an integrating sphere, to fit your LED requirements. For full spectra in the lab see the multi-channel spectro-radiometer, for colour and position checks on assembled boards the LED tester, and for radiometric QA of LED products Radiometry and LED Quality.
Frequently asked questions
QWhat does an integrating sphere measure?
QHow large should an integrating sphere be?
QCan an integrating sphere measure pulsed LEDs and lasers?
Tzachi Sabati
CEO, IZAK Scientific
Physicist specializing in photonics and quantum technologies, with deep expertise in quantum sensors and advanced optical systems. Leads the Advanced Quantum Lab course at the Technion, bridging academic excellence with industry innovation. At IZAK Scientific, provides cutting-edge photonics-based solutions, developing customized inspection and sensing systems for R&D and production. Passionate about advancing quantum sensing applications and integrating novel technologies to meet industry needs.
Get new IZAK guides and tools by email
One short email a month: new technical articles, free calculators and product updates in laser measurement and electro-optics. No spam, unsubscribe anytime.