Profiling CO₂ Laser Beams at 10.6 µm: Attenuation, Profiler Choice and Safety

IZAK Scientific laser beam profiler software analysing a beam profile, shown on a monitor at the IZAK exhibition booth

In short

  • Use an LWIR beam profiler. Detectors for UV–NIR, SWIR or MWIR do not respond at 9.3 or 10.6 µm. You need a profiler built for 7.7–12 µm, and its range must cover your exact line.
  • Sample first, then absorb. One uncoated ZnSe surface reflects about 17 % of a CO₂ beam, so two front-surface pick-offs bring 60 W down to about 1.7 W. Calibrated attenuation does the rest.
  • Watch the polarization. At 45° a ZnSe surface reflects 28 % of s-polarized light but only 8 % of p-polarized light. Work near normal incidence or the reading follows the laser’s polarization.
  • Subtract the room. Objects at room temperature radiate most strongly near 9.7 µm. Take a background frame and let hot parts settle before you measure.
  • It is a Class 4 beam you cannot see. Treat every pick-off, mount and dump as a possible reflection, and use eyewear marked for your wavelength.

Why CO₂ laser beam profiling is different

CO₂ lasers emit in the long-wave infrared, mainly at 10.6 µm and, in many newer marking and cutting sources, at 9.3 µm. Profiling these beams follows the same physics as any other laser: you record the irradiance distribution and calculate second-moment widths, centroid and ellipticity as in ISO 11146. What changes is almost everything around the measurement. A visible or near-IR camera sees nothing at these wavelengths, so you need an LWIR beam profiler, and you have to plan the optics, the attenuation and the safety for a band where your usual habits fail.

Four things set CO₂ beams apart:

  • Optical materials. Ordinary glass and fused silica are opaque at 10.6 µm. Lenses, windows and beam splitters are made of infrared materials such as ZnSe and germanium, which have high refractive indices (about 2.40 and 4.00) and therefore strong surface reflections.
  • Power. Laboratory and industrial CO₂ lasers run from a few watts to several kilowatts. Even a 30 W laser is thousands of times more than a camera sensor can accept.
  • Thermal background. By Wien’s law, a body at 300 K radiates most strongly at 2898 / 300 ≈ 9.7 µm. Your hand, the attenuator housing and the laser head all glow in the same band as the beam.
  • Invisibility. There is no red dot to follow. Alignment relies on low-power targets, the profiler image itself, or a visible guide beam.

Choosing an LWIR beam profiler for 9.3 and 10.6 µm

Start with the wavelength, then match the beam size to the active area. IZAK’s LWIR laser beam profilers come in three configurations. All of them include live 2D/3D profiling, ISO 11146 widths, centroid, ellipticity and orientation, test sequences with pass/fail and reports. The detector response, calibrated attenuation and optics are confirmed for your laser during the engineering review.

ConfigurationSpectral rangeActive widthResolutionCO₂ lines covered
LWIR Compact8–12 µm5.7 mm0.3 MP9.3 and 10.6 µm
LWIR Large-Area8–12 µm12.3 mm1.3 MP9.3 and 10.6 µm
LWIR Wide-Area7.7–10.2 µm15.4 mmApplication dependent9.3 µm only

Two sizing rules decide between them:

  • Keep the beam inside about one third of the active width. ISO 11146 integrates over an area about three times the beam width, and the wings of a real CO₂ mode carry measurable power. A 4 mm beam fits the 12.3 mm sensor, but on the 5.7 mm sensor it would clip the integration area.
  • Put at least ten pixels across the beam. LWIR pixels are several times larger than those of a visible camera, so tightly focused CO₂ spots are hard to profile directly. A 10 mm beam focused by a 100 mm lens gives a waist of roughly 150 µm (4M²λf / πD with M² = 1.1), which is only a handful of LWIR pixels. Measure the focus through a ZnSe relay lens that magnifies it onto the sensor, or measure the collimated beam and calculate the focus. Our guide on measuring a focused laser spot covers the relay method.

Not sure which band or size fits? The laser beam profiler selection guide compares all families from UV to LWIR.

Attenuating a CO₂ beam: sample first, then absorb

A 60 W beam cannot go straight into an absorbing filter in front of a sensor: the filter would have to dissipate almost all of the 60 W, would heat up, change its transmission and re-radiate heat into the image. The robust approach has two stages, as explained in our general guide to beam profiler attenuation:

  1. Sample. Reflect a small, known fraction of the beam from an uncoated surface (a wedge or a pick-off window) and send the transmitted main beam into a water-cooled or air-cooled beam dump.
  2. Attenuate. Bring the sampled beam to the working level of the sensor with calibrated attenuators that only have to handle a watt or two.

The fraction reflected by an uncoated surface follows from the Fresnel equations. Near normal incidence R = ((n − 1) / (n + 1))², which gives 17.0 % for ZnSe (n ≈ 2.40) and 36.0 % for germanium (n ≈ 4.00). Two ZnSe front-surface reflections in series pass 0.170 × 0.170 = 2.9 % of the beam, equal to an optical density of 1.54, without any part of the optics absorbing more than a fraction of a watt.

Pick-off surfaceR at 0–5°R at 45°, s-pol.R at 45°, p-pol.Two surfaces in series, 0–5°
ZnSe, n ≈ 2.4017.0 %28.0 %7.8 %2.9 % (OD 1.54)
Germanium, n ≈ 4.0036.0 %48.4 %23.4 %13.0 % (OD 0.89)

Three practical points follow from the table:

  • Stay close to normal incidence. Most CO₂ lasers are linearly polarized. At 45° the ZnSe reflection changes by a factor of 3.6 between s and p. If the polarization drifts or the laser is rotated, your “calibrated” sample changes with it. At 5° the s and p reflections of ZnSe differ by only about 1 %; at 10° the difference is already 5 %.
  • Use a wedge. The back surface of an uncoated window reflects almost as much as the front. A wedge of a degree or two separates the ghost so that only one reflection reaches the profiler.
  • Let it settle. Pick-offs, dumps and attenuators warm up. Wait until the profile and the total counts are stable before you record results, and keep warm parts out of the profiler’s field of view.
Calculator

CO₂ beam attenuation and sensor-fit calculator

Enter the laser power, the beam diameter at the profiler and the pick-off you plan to use. The calculator gives the peak irradiance of a Gaussian beam, the power after the sampler and in the dump, the optical density still needed to reach your target level at the sensor, and which LWIR configuration holds the beam.

Gaussian beam: peak irradiance I0 = 2P / (πw²) with w = D / 2, twice the mean irradiance inside the 1/e² diameter. Pick-off reflectance from the Fresnel equations for an uncoated surface, n = 2.40 (ZnSe) or 4.00 (Ge), each surface at the same angle and polarization. Required OD = log10(I0 after pick-off / target). The target level here is an example only: the working irradiance for each IZAK configuration is set with the calibrated attenuation in the quotation. Sensor fit uses the one-third rule on the active width.

Worked example: a 60 W CO₂ laser at 10.6 µm

A 60 W CO₂ laser at 10.6 µm has a collimated beam of 4 mm (1/e² diameter) at the planned measurement position. The goal is a production check of beam width, ellipticity and centroid.

Peak irradiance. With w = 2 mm, I0 = 2 × 60 / (π × 0.2²) = 955 W/cm². This is the number that matters for the sensor, not the mean value of 477 W/cm².

Pick-off. Two uncoated ZnSe wedges at 5° reflect 16.96 % each, so 2.87 % of the beam continues: 1.72 W, with a peak irradiance of 27.5 W/cm². The beam dump takes the remaining 58.3 W and needs to be rated, and cooled, for the full laser power.

Attenuation. If the working level at the sensor is 10 mW/cm², the remaining factor is 27.5 / 0.010 = 2,745, an optical density of 3.44. Without the pick-off the same sensor would need OD 4.98 from attenuators that also absorb 60 W.

Sensor fit. 4 mm is 33 % of the 12.3 mm active width of the LWIR Large-Area configuration: a good fit. On the 5.7 mm LWIR Compact the beam would fill 70 % of the width, so either the Large-Area configuration is used or a ZnSe telescope reduces the beam to below about 1.9 mm.

If the wedges are turned to 45°. With a linearly polarized laser the two reflections pass 7.8 % (s) or 0.61 % (p) of the beam instead of 2.9 %. The same set-up would read 2.7 times too bright or 4.7 times too dark, and a polarization change during warm-up would look like a power change. The calculator presets reproduce all of these numbers.

Thermal background, drift and modulated CO₂ lasers

The thermal background is the most common reason for poor LWIR beam profiles. Everything at room temperature emits in the 8–12 µm band, and a warm attenuator housing or the operator’s hand can add a slowly changing offset across the frame. That offset biases second-moment widths upwards, because ISO 11146 weights every pixel by its distance from the centroid squared.

  • Take a background frame with the beam blocked by a room-temperature beam block, at the same exposure, and subtract it. Repeat it after the set-up warms up.
  • Shield the path. A tube or an enclosure between the last attenuator and the sensor keeps warm objects out of the field of view.
  • Apply an aperture or ROI about three times the beam width, as ISO 11146 recommends, so distant background pixels do not enter the calculation.
  • Mind modulation. Many RF-excited CO₂ lasers set their power by pulse-width modulation at kilohertz rates, and TEA or Q-switched lasers emit discrete pulses. LWIR array sensors respond more slowly and run at lower frame rates than visible cameras, so plan the exposure to average many modulation periods, and report average power, not peak power, in the calculator. For true single-pulse work, see our guide to profiling pulsed laser beams.

Safety at 10.6 µm

Any CO₂ laser above 0.5 W CW is Class 4 under IEC 60825-1. The beam is invisible and is absorbed in the front of the eye and in the skin, so the main injuries are corneal and skin burns, and the beam can ignite paper, plastics and fabrics. For exposures longer than 10 s, the maximum permissible exposure at this wavelength is about 0.1 W/cm² (1000 W/m²). The 60 W beam in the example exceeds that by almost four orders of magnitude.

  • Every pick-off makes a second beam. The transmitted 58 W and the ghost reflections must end in dumps. Enclose the sampling optics where possible.
  • Metal mounts and tools reflect. Polished or anodised surfaces that look matt in visible light can be good mirrors at 10.6 µm.
  • Eyewear must be marked for your wavelength. Check the optical density label at 10.6 or 9.3 µm; the colour of the lens tells you nothing in the infrared.
  • Align at low power. Use the laser’s lowest stable setting or a visible guide beam, then raise the power with the profiler already behind the attenuation.
  • Handle ZnSe carefully. Use gloves, and treat cracked or burnt ZnSe optics as hazardous waste.

Common CO₂ beam profiling mistakes

Using the wrong bandno signal

A 9.3 µm laser fits all three LWIR configurations, but a 10.6 µm laser needs an 8–12 µm range. Check the exact line before you order.

Absorbing the full beam in front of the sensorheat, drift

Absorbing filters that take tens of watts heat up, change transmission and glow in the image. Sample the beam first and dump the rest.

Pick-off at 45° with a polarized laserfalse power change

The sampled fraction then depends on the polarization. Work near normal incidence or calibrate for the actual polarization and keep it fixed.

No background subtractionwidths too large

The room-temperature background adds to the wings of the profile and inflates D4σ. Subtract a blocked-beam frame and use an ROI.

Profiling a tight focus directlytoo few pixels

A 150 µm CO₂ focus covers only a few LWIR pixels. Use a relay lens or measure the collimated beam and calculate the focus.

IZAK products for this application

IZAK LWIR beam profilers cover 8–12 µm for 9.3 and 10.6 µm CO₂ lasers, with a 7.7–10.2 µm wide-area option for 9.3 µm. Each system is supplied with the attenuation and optics defined for your laser, and runs the same IZAK software as our other profilers: ISO 11146 widths, centroid, ellipticity, test sequences, pass/fail and reports.

Need a complete CO₂ beam test station with sampling optics, beam dump and automated reporting? See custom electro-optical test systems or book a beam profiler consultation.

Profiling a CO₂ laser?Send the wavelength (9.3 or 10.6 µm), power, beam diameter at the measurement plane and whether the laser is CW, modulated or pulsed. We will recommend the LWIR configuration, sampling and attenuation.
Contact an engineer

Frequently asked questions

QCan a standard beam profiler measure a CO₂ laser at 10.6 µm?
No. Profilers for UV–NIR, SWIR or MWIR do not respond at 10.6 µm. You need an LWIR beam profiler whose spectral range covers your line: 8–12 µm for 10.6 µm, while 9.3 µm lasers can also use a 7.7–10.2 µm configuration.
QHow much attenuation does a CO₂ beam profiler need?
It depends on the peak irradiance, not only the power. A 60 W beam of 4 mm diameter has a peak of about 955 W/cm². Two uncoated ZnSe pick-offs near normal incidence reduce that to about 27 W/cm² (OD 1.54), and calibrated attenuators provide the rest, about OD 3.4 for a 10 mW/cm² working level.
QWhy does my LWIR beam image show a background or drift?
Objects at room temperature radiate most strongly near 9.7 µm, in the same band as the beam. Warm attenuators, dumps or hands add an offset that inflates the measured width. Subtract a blocked-beam background frame, shield the beam path, use an ROI about three times the beam width and let the set-up reach thermal equilibrium.

References

  1. ISO 11146-1:2021, Lasers and laser-related equipment — Test methods for laser beam widths, divergence angles and beam propagation ratios — Part 1: Stigmatic and simple astigmatic beams.
  2. IEC 60825-1:2014, Safety of laser products — Part 1: Equipment classification and requirements.
  3. ISO 11554:2017, Optics and photonics — Lasers and laser-related equipment — Test methods for laser beam power, energy and temporal characteristics.
>Tzachi Sabati Tzachi Sabati
CEO, IZAK Scientific
Physicist specializing in photonics and quantum technologies, with deep expertise in quantum sensors and advanced optical systems. Leads the Advanced Quantum Lab course at the Technion, bridging academic excellence with industry innovation. At IZAK Scientific, provides cutting-edge photonics-based solutions, developing customized inspection and sensing systems for R&D and production. Passionate about advancing quantum sensing applications and integrating novel technologies to meet industry needs.

 

Get new IZAK guides and tools by email

One short email a month: new technical articles, free calculators and product updates in laser measurement and electro-optics. No spam, unsubscribe anytime.

Get new IZAK guides and free calculators by email, one short email a month: Subscribe · Privacy Policy
Skip to content